Pamaraters
1. 128*64 HD LCD Display, high capacity words, friendly user interface, dynamically display working status;
2. Female Voice alarm in English/Russian (other language is optional for customized machine)
There are bubbles,Tube is blocked,Close the door,IV set Mismatched,Battery running out,Infusion finish,Equipment failure ,Abnormally Operation,Software Exception,Low battery,forget the operations etc.
3. Compatible with any brands of infusion sets after correct calibration;
4. Preset solution volume to greatly reduce the workload of nurses;
5. Work mode: ml/h and drop/min can switch freely;
6. Three levels of occlusion: high, middle and low;
7. Purge /Bolus function;
8. KVO (keep-vein-open) automatically opens as infusion is completed, KVO rate is 1ml/h---5ml/h (1ml/h step);
9. Power Source:100-240Vac, 50/60Hz;Internal rechargeable Li battery, capacity≥1,800mAh, voltage=11.1V, 4 hours backup time at the rate of 25ml/h (appointed medium rate in IEC60601-2-24) for the new battery charged for 12 hours;
10. Automatically record the settings of last infusion;
11. 2000pcs of infusion records and has drug library with 1000 durgs.
Infusion Rate | ml/h mode: 0.1~1200ml/h (1ml/h step) |
drop/min mode: 1~266drop/min (1drop/min step) | |
Purge Rate | 100ml/h---1200ml/h (can be calibrated) |
Bolus | 800ml/h |
Flow Rate Accuracy | ml/h mode: Within ±5% (after correct calibration) |
drop/min mode: Within ±3% (after correct calibration) | |
Mechanical Precision | Within ±2% |
Maximum infusion pressure | 150kPa |
Volume Limit | 1---9999ml |
Total Infusion Volume | 0---9999.9ml |
Alarm Pressure | High 800mmHg±200mmHg (106.7kPa±26.7kPa) |
Medium 500mmHg±100mmHg (66.7kPa±13.3kPa) | |
Low 300mmHg±100mmHg (40.7kPa±13.3kPa) | |
Audible and Visible Alarms | There are bubbles,Tube is blocked,Close the door,IV set Mismatched,Battery running out,Infusion finish,Equipment failure ,Abnormally Operation,Software Exception,Low battery,forget the operations etc |
Power Source | 100-240Vac, 50/60Hz; Internal rechargeable Li battery, capacity≥1,800mAh, voltage=11.1V, 4 hours backup time at the rate of 25ml/h (appointed medium rate in IEC60601-2-24) for the new battery charged for 12hours. |
Bubble Detector | Ultrasonic wave detector; detection sensitivity ≥25μL |
KVO Speed | 1---5ml/h (1ml/h step) |
Power Consumption | 40VA |
Fuse | T3.15AH/250V, 2pcs installed |
IP Classification | IPX2 |
Equipment Classification | Class I, internal power supply, Type CF |
Operation Environment | Temperature: +5~+40 degree |
Relative Humidity: 20~90% | |
Transport & Storage Condition | Ambient Temperature:-20℃~+55℃ |
Relative Humidity:≤95% | |
Atmospheric Pressure: 50.0KPa~106.0KPa | |
Infusion Set | Compatible with any brand of infusion sets after correct calibration |
Infusion maximum value(under single fault) | 1ml |
Forbidden | Not support blood Infusion |
Application field | internal medicine, surgical, pediatrics, obstetrics and gynecology, ICU, CCU, operation room and other clinical infusion treatment,etc.Also support the veterinary purpose |
Dimension | 140mm(L) x 157mm(W) x 220mm(H) |
Net Weight | 1.8kg |
Package Reference | 1pc Carton: 2.8KG, 30cm(L) x 21cm(W) x 24cm(H) |
2pcs Carton: 5.6KG, 45cm(L) x 25cm(W) x 33cm(H) | |
3pcs Carton:8.4KG, 67cm(L) x 25cm(W) x 33cm(H) |